Click Click here

You are here: Using the DHR/AR > Calibrating the DHR/AR > DHR/AR Calibration Overview

DHR/AR Rheometer Calibration Overview

To ensure accurate experimental results, calibrate the rheometer upon initial installation. For best results, recalibrate the rheometer periodically. Note that calibration procedures vary. The table below lists required calibration/verification tasks and the recommended interval for each, as well as when to perform calibrations following a transducer replacement.

For information on temperature systems calibration, refer to DHR/AR Temperature Calibration.

 

Calibration Tasks and Recommended Intervals for AR Rheometer

Calibration Task

Calibration Interval

Instrument Inertia

Suggested: At least every 3 months.

Rotational Mapping

Suggested: As the torques in your procedure drop, the error from an unmapped bearing will increase, so it is necessary to perform mapping to keep these errors to a minimum.

For the AR Series: For maximum accuracy at the lower torque end of the rheometer, it is recommended that the mapping be performed each time a new measuring system is used, or if the current geometry is removed for cleaning.

For the DHR Series: Provided that the geometry is refitted in the same position (see Fitting a Geometry on the DHR) and the geometry has previously been mapped, the re-mapping is not necessarily required unless low torque or creep recovery tests are being performed.

 

Oscillatory Mapping

Suggested: Use oscillatory mapping to get the best data at low torque, low displacement, and especially low torque and low displacement.

Axial Mapping Suggested: Every time the instrument is set up to run DMA experiments.

© 2014 TA Instruments–Waters LLC | 159 Lukens Drive | New Castle, DE 19720